Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

This recall has been terminated (originally issued January 27, 2017).

Company
Immuno-Mycologics, Inc
Recall Initiated
December 27, 2016
Posted
January 27, 2017
Terminated
March 8, 2017
Recall Number
Z-1109-2017
Quantity
630 units
Firm Location
Norman, OK
Official Source
View on FDA website ↗

Reason for Recall

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

Distribution

Nationwide throughout the US

Lot / Code Info

161028133653JH, 161003111544KH, and 161007105036JG

Root Cause

Material/Component Contamination

Action Taken

Immuno sent an Urgent: Medical Device Recall Letter dated December 27, 2016, to all affected customers. All customers were notified via emails to stop using the product on December 27th and December 28th. A recall notification and response form was attached to the email. For further questions, please call (405)364-1058.