AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage: TomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications

This recall has been terminated (originally issued February 10, 2020).

Company
Agfa N.V. Septestraat 27 Mortsel Belgium
Recall Initiated
February 10, 2020
Terminated
March 30, 2023
Recall Number
Z-1502-2020
Quantity
2 units
Official Source
View on FDA website ↗

Reason for Recall

Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.

Distribution

US Nationwide distribution in the states of FL, VA.

Lot / Code Info

Serial Numbers: 00949, 1161

Root Cause

Under Investigation by firm

Action Taken

AFGA issued via email On February 10, 2020, an URGENT FIELD SAFETY NOTICE letter. Letter states reason for recall, health risk, and action to take: With this letter, Agfa is informing you that the DR 800 system still can be used including for Tomosynthesis for mAs values which are larger than 0,2 mAs. In case Tomosynthesis image acquisition sequence does not stop automatically after the expected number of exposures, release the exposure button as this will stop the sequence and notify your Agfa Service contact at once. Contact your local Agfa Imaging organization 1-877-777-2432 and please reference Problem Record PRB2000387 and Vigilance Record VR0000716. Complete/Return by fax or email the URGENT FIELD SAFETY NOTICE Feedback form