DP-30 Digital Ultrasonic Diagnostic Imaging System, Part No. 1155E-PA00004 - Product Usage: The Digital Ultrasonic Diagnostic Imaging System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, muscular-skeletal (conventional, superficial), cardiac (adult, pediatric), peripheral vascular and urology exams.

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
January 13, 2020
Posted
February 5, 2020
Terminated
August 21, 2020
Recall Number
Z-1330-2020
Quantity
53 DP-30 systems and 18 USB's containing the affected software
Firm Location
Mahwah, NJ

Reason for Recall

The DP-30 displays an incorrect needle-guide bracket when used with the 65EC10EA model transducer.

Distribution

US Nationwide distribution in the states of CA, GA, IL, MD, NY, SC, and WA.

Lot / Code Info

DP-30 Ultrasound with software version 03.00.00 and 03.01.00

Root Cause

Software Design Change

Action Taken

On January 13, 2020, the firm distributed Urgent Medical Device Correction letters to affected distributors. Distributors were informed that the system displays the incorrect needle-guide bracket when used with the 65EC10EA model transducer. Customers were advised to avoid use of NGB-004 needle-guide with transducer model 65EC10EA until a software update has been applied to the system. The system may be used normally for all other functions. The firm recommended that distributors immediately notify any clients to which they have distributed affected systems (DP-30 systems with software version 03.00.00 and 03.01.00 distributed between 06/25/2018 and 06/28/2019), or provided a software update (USB with V03.01.00, part number 110-002261-05). Within the next several weeks, Mindray will provide a software update at no cost. Should you require technical assistance while performing the update, please contact Mindrays Technical Support team at 800-288-2121, option 2, Monday through Friday, 8:30am - 5:30pm PT. The replacement USB drive contains instructions on how to perform the software update, under the Manuals folder in the DP-10 & DP-20 & DP-30 Series System Recovery Guide file. If you have questions regarding this recall, you may contact Diane Arpino, Director, Quality Operations and Regulatory Affairs, at 201-995-8407, or via email at d.arpino@mindray.com.