RecallDepth

MicroScan LabPro Information Manager System, Software Versions 1.0 to 4.41 Catalog No. 10638819, 10638820, 10638823, 10638824, 10638825,10638826, 10714149,10714159, 10805072, 10805073, 10975000, 10975001, 6000-0008, 6000-0026. Japan: 10638819 LabPro v4.11 Software Update Kit 10638820 LabPro v4.11 System Software 10805072 LabPro MBT Non-Connect Systems v4.30 10805073 LabPro MBT Connect Systems v4.30 Canada: 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit Europe: 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software USA: 10714149, LabPro v4.11 Software Update Kit 10714150 LabPro v4.11 System Software 10975000 ASM LabPro-MBT v4.40 10975001 ASM LabPro-MBT v4.40 for Connect 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit ROW: 6000-0008 LabPro v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software Product Usage: LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel. LabPro AlertEx is a functional subset of the LabPro Data Management System that analyzes MicroScan ID and AST data, or other predefined parameters, against a series of established rules/alerts and notifies the user of unusual and/or critical conditions, which may warrant further analysis or actions.

Company
Beckman Coulter Inc.
Recall Initiated
February 11, 2016
Terminated
January 4, 2018
Recall Number
Z-1269-2016
Quantity
2,702 units total (1,032 units in US)
Firm Location
Brea, CA

Reason for Recall

Beckman Coulter is recalling the MicroScan LabPro Information Manager System because the software incorrectly allows the operator to manually edit the carbohydrate substrates when manually reading dried overnight gram negative panels with an ID Hold status.

Distribution

Affected geographies: Bahamas, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Lithuania, Mexico, Morocco, Netherlands, Northern Mariana Islands, Panama, Paraguay, Peru, Poland, Portugal, Puerto Rico, Romania, Russia, Singapore, South Africa, Spain, Taiwan, Tunisia, Turkey, United States, Uruguay, Virgin Islands, Zimbabwe.

Lot / Code Info

not applicable

Root Cause

Unknown/Undetermined by firm

Action Taken

An Urgent Medical Device Recall Letter dated 2/11/2016 was sent to all customers who purchased the MicroScan LabPro Information Manager System, Version 1.0 to Version 4.41. Beckman Coulter is expanding the recall that was initiated in July 2015 regarding FA-25776 for the MicroScan LabPro Information Manager System. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Representative: http:www.beckmancoulter.com, (800) 677-7226 in US and Canada. Customers outside of US and Canada, are instructed to contact their local Beckman Coulter representative.

More recalls by Beckman Coulter Inc.

View all recalls by this company →