A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.
This recall has been terminated (originally issued December 18, 2015).
- Company
- Biomet, Inc.
- Recall Initiated
- November 20, 2015
- Posted
- December 18, 2015
- Terminated
- September 8, 2016
- Recall Number
- Z-0450-2016
- Quantity
- 182, 965
- Firm Location
- Warsaw, IN
- Official Source
- View on FDA website ↗
Reason for Recall
These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.
Distribution
AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR CHILE, COSTA RICA, NETHERLANDS, CANADA
Lot / Code Info
Part number and size (non-sterile) 816140010, 4.0X10MM; 816140012, 4.0X12MM; 816140014, 4.0X14MM; 816140016, 4.0X16MM; 816140018, 4.0X18MM; 816140020, 4.0X20MM; 816140022, 4.0X22MM; 816140024, 4.0X24MM; 816140026, 4.0X26MM; 816140028, 4.0X28MM; 816140030, 4.0X30MM; 816140032, 4.0X32MM; 816140034, 4.0X34MM; 816140036, 4.0X36MM; 816140038, 4.0X38MM; 816140040, 4.0X40MM; 816140042, 4.0X42MM; 816140044, 4.0X44MM; 816140046, 4.0X46MM; 816140048, 4.0X48MM; 816140050, 4.0X50MM; 816140055, 4.0X55MM; 816140060, 4.0X60MM; 816140065, 4.0X65MM; 816140070, 4.0X70MM; 816140075, 4.0X75MM; 816140080, 4.0X80MM; 816140085, 4.0X85MM; 816140090, 4.0X90MM; 816140095, 4.0X95MM; 816140100, 4.0X100MM Part number and size (sterile) 856140010, 4.0X10MM; 856140012, 4.0X12MM; 856140014, 4.0X14MM; 856140016, 4.0X16MM; 856140018, 4.0X18MM; 856140020, 4.0X20MM; 856140022, 4.0X22MM; 856140024, 4.0X24MM; 856140026, 4.0X26MM; 856140028, 4.0X28MM; 856140030, 4.0X30MM; 856140032, 4.0X32MM; 856140034, 4.0X34MM; 856140036, 4.0X36MM; 856140038, 4.0X38MM; 856140040, 4.0X40MM; 856140042, 4.0X42MM; 856140044, 4.0X44MM; 856140046, 4.0X46MM; 856140048, 4.0X48MM; 856140050, 4.0X50MM; 856140055, 4.0X55MM; 856140060, 4.0X60MM; 856140065, 4.0X65MM; 856140070, 4.0X70MM. All lots manufactured between 06/01/08 and 06/01/15.
Root Cause
Process change control
Action Taken
On 11/20/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors, hospital risk managers, and implanting surgeons via courier. Distributors were instructed to locate and return the affected product, hospitals to locate and discontinue use, and implanting surgeons to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.