RecallDepth

DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Expert Lateral Femur Nail System and indicated to stabilize Subtrochanteric fractures, Ipsilateral neck/shaft fractures, Femoral shaft fractures, Impending pathologic fractures, and malunions/Non-unions.

This recall has been terminated (originally issued June 26, 2015).

Company
Synthes, Inc.
Recall Initiated
May 27, 2015
Posted
June 26, 2015
Terminated
December 18, 2015
Recall Number
Z-1915-2015
Quantity
276
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

certain lots of the Titanium Recon Screws (6.5mm ) were found to contain the incorrect insert GP0760-A (Synthes Titanium Buttress Locking Plate System), instead of the correct insert GP0706-E (Synthes Titanium Intramedullary Nails and Components).

Distribution

Nationwide Distribution.

Lot / Code Info

Part Numbers :  04.003.024 04.003.025 04.003.026 04.003.027 04.003.028 04.003.029 04.003.031  with Lot Numbers:  8638958 8638957 8534954 8638955 8556020 8556018 8653032 8653036

Root Cause

Process control

Action Taken

DePuy Synthes sent an Urgent Medical Device Safety Notification, dated May 27, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions Required: Please take the following actions: Review the information listed in this notification. Forward this notice to anyone in your facility that needs to be informed. Complete the attached Verification Section (page 3 of this notification). Please include your name, title, address, telephone number and signature in the spaces provided. Send a copy of the completed Verification Section by: Fax: (888) 912-2182 or Scan/email: Synthes4844@stericycle.com If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page three (3) of this notification. Maintain a copy of this notice. If you have any questions, please call (610)719-5450. Thank you for your attention to this issue.