GE Healthcare Innova IGS 520. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human anatomy for diagnostic, interventional and image-guided surgical procedures.

Company
Ge Healthcare
Recall Initiated
May 29, 2015
Posted
July 3, 2015
Terminated
October 28, 2015
Recall Number
Z-1980-2015
Quantity
10
Firm Location
Waukesha, WI

Reason for Recall

GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).

Distribution

Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, GA, IL, IN, KS, MS, NV, OH, OK, OR, PA, TN, TX, VA, WA and WI; and countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAPAN, RUSSIA, UNITED KINGDOM, and VIETNAM.

Lot / Code Info

Serial Number System ID 0000013C2S0065 302421IGS520 15C2R0058 LVIGS2 15C2R0065 AM1396F508 15C2R0082 206598UIGS 15C2R0092 AM135FE37 0000015C2R0062 HU1117VA01 0000015C2R0057 083016848125215 0000015C2R0069 5512XR0026  0000015C2R0060 00632VAS01 0000015C2R0080 00391VAS02

Root Cause

Nonconforming Material/Component

Action Taken

The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter on 5/29/2015- GEHC Ref#12231- dated May 29, 2015 to their Consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and contact Information. The customers were instructed that you may continue to use the system and avoid rapid direction changes with the gantry motions and prevent potential collisions with other equipment. If the failure occurs during interventional examination, please ensure that you have established procedures for handling patients in case of the loss of gantry motion as stated in your product labeling. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for this correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.