Bone Marrow Aspiration Needle 15G x 2.688 MAX, Part Number DBMNI1501; Bone Marrow Aspiration Needle 15G x 4 MAX, Part Number DBMNI1504; Bone Marrow Aspiration Needle 16G x 2.688 MAX, Part Number DBMNI1601

This recall has been terminated (originally issued July 21, 2015).

Company
Argon Medical Devices, Inc
Recall Initiated
June 18, 2015
Posted
July 21, 2015
Terminated
June 23, 2016
Recall Number
Z-2188-2015
Quantity
6390 units
Firm Location
Athens, TX
Official Source
View on FDA website ↗

Reason for Recall

Potential defect in the forming of the primary package of the blister that may compromise the sterility of the product inside the tray.

Distribution

US

Lot / Code Info

Lot Numbers: 11088595, 11089104, 11089369, 11089370, 11094673, 11094679, 11095843, 11096115, 11098022, 11098023, 11098643, 11098707

Root Cause

Packaging process control

Action Taken

The recalling firm's sales force was notified of the recall on 6/17/15. The recalling firm will begin to send out Urgent Medical Device Recall letters by registered mail to affected customers on 6/18/15. Customers are instructed to inspect and quarantine and return any affected product and complete the Recal Response Form.