The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or muscles when connected to an extension and/or lead. The Medtronic InterStim II Neurostimulator, Model 3058, and Medtronic InterStim Neurostimulator, Model 3023, are parts of neurostimulation systems for InterStim Therapy. The neurostimulators are multiprogrammable devices that accommodates a lead through which a stimulation program is delivered. The Medtronic Enterra Neurostimulator, Model 3116, is part of an Enterra Therapy system for gastric electrical stimulation (GES). The neurostimulator is a multiprogrammable device that accommodates a lead through which a stimulation program is delivered.

This recall has been terminated (originally issued June 26, 2015).

Company
Medtronic Neuromodulation
Recall Initiated
January 9, 2015
Posted
June 26, 2015
Terminated
September 28, 2017
Recall Number
Z-1147-2015
Quantity
396,975 (258,805 USA, 138,170 OUS)
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.

Distribution

Worldwide Distributon- USA( nationwide including DC and PR) and Internationally to Mexico and Canada

Lot / Code Info

All lot/serial numbers are affected for all 4 Model numbers:  " Model 7425  " Model 3058  " Model 3023  " Model 3116

Root Cause

Other

Action Taken

Consignees were sent a Medtronic {Medical Device Correction" letter dated January 9, 2015. The letter was addressed to Healthcare Professional. The letter described the issue and the product involved in the the recall. The letter also listed the Occurrence of the problem and labeling update. A Physician Reply Form was included which requested consignees to return the form to Medtronics. For questions they can contact Medtronic Neuromodulation Technical Services at 1-8007070944 weekdays 7 am 6 pm Central time. A second "Physician Notification Confirmation Form" was included to be completed by Medtronic Representatives.