ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Model # ATS45. Manfuactured by Ethicon Endo-Survery, LLC; 475 Calle C; Guaynabo PR 00969.

This recall has been terminated (originally issued November 15, 2010).

Company
Ethicon Endo-Surgery Inc
Recall Initiated
January 13, 2009
Posted
November 15, 2010
Terminated
May 10, 2012
Recall Number
Z-0361-2011
Quantity
4,113 UNITS
Firm Location
Cincinnati, OH
Official Source
View on FDA website ↗

Reason for Recall

The mechanism that connects the articulation joint to the device shaft may cause the jaws of the cutters to remain closed and clamped down on tissue after the device is fired. If the jaws of the device remain clamped down on the tissue, there is a risk the surgery will need to be modified to remove the device. This may lead to an increase in procedure time, tissue manipulation, and a change in t

Distribution

Worldwide Distribution: USA only, in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI. IA. ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.

Lot / Code Info

Lot #'s: E4MD1T, E4MD5W, E4MD3Y, E4ME7J, E4MD4Z, E4ME7U, E4MD50 & E4MF0A

Root Cause

Other

Action Taken

Ethicon Endo-Surgery issued "Urgent Device Recall Event 1952" notification letters dated January 21, 2009 to their customers. They instructed their custormers to immediately discontinue use of the product. They are to complete and fax back the Business Reply Form. Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle. Replacement product should be received within 30-60 days after return. Customers can contact Stericycle concerning this recall at 1-866-861-6921, reference Event 1952.