RecallDepth

Alto DR (model 614) and Alto 2 DR (model 624) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Dual chamber arrhythmia detection. Antibradycardia pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring atrial and ventricular activity to identify persistent ventricular arrhythmias and to deliver appropriate therapies.

Company
Ela Medical Llc
Recall Initiated
July 19, 2005
Posted
September 24, 2005
Terminated
January 28, 2007
Recall Number
Z-1587-05
Quantity
392 total devices (23-model 614 & 369-model 624)
Firm Location
Plymouth, MN

Reason for Recall

Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.

Distribution

Nationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI.

Lot / Code Info

devices manufactured between April 2003 and August 2004. Model 614 serial numbers: 218XF003, 218XF012, 218XF014, 218XF022, 218XF042, 220XF012, 220XF014, 220XF017, 220XF018, 220XF019, 220XF020, 220XF030, 220XF038, 220XF061, 220XF062, 220XF064, 221XF018, 221XF042, 222XF034, 222XF074, 223XF057, 223XF059 and 223XF061.   Model 624 serial numbers: 238XJ013, 238XJ023, 238XJ037,  244XJ044, 244XJ048, 244XJ062, 244XJ076, 244XJ079,  244XJ092, 246XJ022, 246XJ029, 246XJ048, 246XJ073,  246XJ081, 249XJ028, 249XJ048, 249XJ088, 249XJ160,  249XJ168, 304XJ002, 304XJ018, 304XJ023, 304XJ035,  304XJ041, 304XJ063, 304XJ072, 304XJ081, 304XJ086,  304XJ096, 304XJ100, 304XJ109, 304XJ113, 307XJ015, 307XJ049, 307XJ052, 307XJ057, 307XJ065, 307XJ124,  307XJ142, 307XJ153, 307XJ162, 307XJ171, 307XJ176,  307XJ220, 307XJ234, 307XJ241, 307XJ256, 313XJ013,  313XJ014, 313XJ025, 313XJ027, 313XJ033, 313XJ036, 313XJ038, 313XJ044, 313XJ046, 313XJ068, 313XJ071,  313XJ072, 313XJ081, 313XJ086, 313XJ088, 313XJ090,  313XJ092, 313XJ098, 313XJ104, 313XJ122, 313XJ134,  313XJ136, 313XJ137, 313XJ151, 313XJ158, 313XJ167,  313XJ167, 313XJ170, 313XJ178, 313XJ183, 313XJ191,  313XJ196, 313XJ198, 313XJ199, 313XJ205, 313XJ206,  313XJ210, 313XJ214, 313XJ215, 313XJ217, 313XJ224, 313XJ235, 313XJ235, 313XJ237, 313XJ242, 313XJ242,  313XJ245, 313XJ249, 313XJ250, 313XJ270, 313XJ275,  314XJ003, 314XJ006, 314XJ097, 314XJ102, 314XJ109,  314XJ134, 314XJ136, 314XJ139, 314XJ140, 314XJ148,  314XJ157, 314XJ158, 314XJ160, 314XJ174, 314XJ186,  314XJ187, 314XJ198, 314XJ212, 314XJ230, 314XJ239,  314XJ254, 314XJ256, 314XJ295, 339XJ012, 339XJ015,  339XJ016, 339XJ017, 339XJ019, 339XJ020, 339XJ023,  339XJ031, 339XJ034, 339XJ034, 339XJ037, 339XJ039,  339XJ041, 339XJ042, 339XJ048, 339XJ049, 339XJ050,  339XJ051, 339XJ053, 339XJ055, 339XJ057, 339XJ057,  340XJ001, 340XJ013, 340XJ014, 340XJ017, 340XJ024,  340XJ025, 340XJ030, 340XJ034, 340XJ040, 340XJ044,  340XJ045, 340XJ047, 340XJ051, 340XJ053, 340XJ057, 340XJ061, 340XJ062, 340XJ067, 340XJ076, 340XJ091,  340XJ092, 340XJ098, 340XJ106, 340XJ108, 340XJ117,  342XJ005, 342XJ006, 342XJ009, 342XJ016, 342XJ027,  345XJ002, 345XJ038, 345XJ048, 345XJ049, 345XJ050,  345XJ053, 345XJ064, 345XJ071, 349XJ001, 349XJ013,  349XJ014, 349XJ019, 349XJ020, 349XJ041, 350XJ010,  350XJ024, 350XJ025, 350XJ027, 350XJ030, 350XJ034,  350XJ040, 350XJ043, 350XJ056, 350XJ057, 350XJ060,  350XJ076, 350XJ076, 350XJ077, 350XJ080, 350XJ083,  351XJ003, 351XJ005, 351XJ007, 351XJ014, 351XJ018, 351XJ019, 351XJ020, 351XJ027, 351XJ029, 351XJ038,  351XJ067, 351XJ069, 351XJ075, 351XJ078, 351XJ082,  402XJ005, 402XJ018, 402XJ022, 402XJ023, 402XJ024,  402XJ026, 402XJ045, 402XJ053, 402XJ060, 402XJ061,  402XJ066, 402XJ074, 402XJ076, 402XJ078, 402XJ080,  402XJ085, 402XJ086, 402XJ087, 402XJ088, 403XJ007,  403XJ008, 403XJ017, 403XJ018, 403XJ026, 403XJ027,  403XJ031, 403XJ034, 403XJ040, 403XJ048, 403XJ050,  403XJ054, 403XJ060, 403XJ073, 403XJ086, 403XJ093,  403XJ095, 403XJ098, 403XJ100, 403XJ104, 403XJ105,  403XJ107, 405XJ003, 405XJ013, 405XJ030, 405XJ049,  405XJ053, 405XJ056, 405XJ067, 405XJ070, 406XJ031,  406XJ033, 406XJ035, 406XJ036, 406XJ039, 406XJ040, 406XJ044, 406XJ065, 406XJ069, 406XJ069, 406XJ070,  406XJ075, 407XJ035, 409XJ007, 409XJ009, 409XJ012,  409XJ017, 409XJ018, 409XJ020, 409XJ032, 409XJ034,  409XJ035, 409XJ037, 409XJ041, 409XJ043, 409XJ044,  409XJ046, 409XJ049, 409XJ052, 409XJ053, 409XJ054,  409XJ059, 410XJ004, 410XJ022, 410XJ026, 410XJ028,  410XJ030, 411XJ025, 411XJ035, 411XJ036, 411XJ042,  411XJ045, 411XJ049, 411XJ050, 411XJ054, 411XJ055,  411XJ057, 413XJ001, 413XJ004, 413XJ005, 413XJ006,  413XJ007, 413XJ009, 413XJ016, 413XJ017, 413XJ032,  413XJ040, 413XJ044, 413XJ049, 413XJ050, 413XJ051,  413XJ052, 413XJ054, 413XJ056, 413XJ061, 413XJ062,  413XJ067, 414XJ005, 414XJ009, 414XJ017, 414XJ037,  414XJ038, 414XJ039, 414XJ042, 414XJ047, 414XJ050,  414XJ063, 414XJ070, 414XJ074, 414XJ085, 414XJ093,  414XJ101, 414XJ102, 414XJ105, 414XJ107, 415XJ005,  415XJ022, 415XJ024, 415XJ033, 415XJ035, 415XJ037,  415XJ046, 415XJ068, 415XJ071, 415XJ094, 416XJ021,  416XJ035, 416XJ071, 417XJ057, 417XJ059, 417XJ064 and 418XJ010.

Root Cause

Other

Action Taken

A Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician.