Medtronic MiniMed Paradigm Model 712
- Company
- Medtronic Minimed
- Recall Initiated
- September 14, 2004
- Terminated
- March 3, 2005
- Recall Number
- Z-0010-05
- Quantity
- 10,600
- Firm Location
- Northridge, CA
Reason for Recall
Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The investigation revealed this only occurs when the pump is programmed to display information in Spanish.
Distribution
Nationwide & Spain
Lot / Code Info
Paradigm model 712pumps with serial numbers that end in '-Dxx3' . Where 'x' can be any number or letter.
Root Cause
Other