Quanta Technologies L.L.C. brand REORDER # LAPKIT 3. Contents 1 -55060 Suction/Irrigation Valve 2- 30012 Insufflation Tubing set 3- 5010 Anti Fog Solution with Sponge. Manufactured for Quanta Technologies, L.L.C., Tucson, AZ.
- Company
- Vital Concepts, Inc.
- Recall Initiated
- July 11, 2003
- Posted
- October 15, 2003
- Terminated
- April 12, 2004
- Recall Number
- Z-0024-04
- Quantity
- 160
- Firm Location
- Grand Rapids, MI
Reason for Recall
The suction valve may become stuck in the 'suction on' position under vacuum.
Distribution
Arizona, Illinois, Indiana, Michigan, Ohio, Pennsylvania, South Carolina, Texas, Utah, Washington, Canada, Israel and Japan
Lot / Code Info
los 994985, 994986 and 995044.
Root Cause
Other
Action Taken
A request for product return was sent to each customer beginning on July 11, 2003.
More recalls by Vital Concepts, Inc.
Reorder # Lapkit 3. Contents: 1- 55060 suction/irrigation valve 2- 30012 insufflation tubing set...
Jul 20, 2004
Auto Suture IRRIVAC MAX Suction irrigation sytem with 5 mm, 33 cm probe with double spike tubing. S...
Jul 20, 2004
Irrigation aspiration system for Nexhat Dorsey type pumps with 33 cm probe and tubing, latex free, c...