RecallDepth

One Way Valve/Male to Female Flow, 22 mm I.D./22 mm O.D., Ref 396799, Lot ***, SINGLE USE ONLY, Latex Free, Quantity 50 Each Manufactured for and Distributed by RUSCH 2450 Meadowbrook Parkway, Duluth, GA 30098, Made in Taiwan.

This recall has been terminated (originally issued July 20, 2004).

Company
Rusch Inc
Recall Initiated
February 3, 2004
Posted
July 20, 2004
Terminated
March 21, 2006
Recall Number
Z-0580-04
Quantity
13650 pieces
Firm Location
Duluth, GA
Official Source
View on FDA website ↗

Reason for Recall

Valve cap may become loose and enter the air circuitry.

Distribution

Nationwide and Canada

Lot / Code Info

Catalog No. 396799; Lot numbers: 010905, 011020, 020305, 020530, 020705, 020925, 030120, 031120, 030325, 030415.

Root Cause

Other

Action Taken

Consignees were notified on 2/3/2004 via certified mail and telephone.