GE Healthcare Proteus XR/a Eclipse Collimator model # 2379827.

This recall has been terminated (originally issued September 30, 2008).

Company
Ge Healthcare
Recall Initiated
March 29, 2007
Posted
September 30, 2008
Terminated
February 4, 2012
Recall Number
Z-1713-2008
Quantity
132
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

The actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.

Distribution

Worldwide Distribution --- USA including states of CA, DE, FL, GA, ID, IL, KS, KY, MD, MI, MO, MT, NC, NJ, NM, NY, OH, PA, SC, TX, UT, VA, and WV, and countries of Australia, Canada, China, France, Germany, Greece, Italy, Korea, Poland, Saudi Arabia, United Kingdom, and Venezuela.

Lot / Code Info

42948HL0, 42949HL8, 43054HL6, 43055HL3, 43056HL1, 43057HL9, 43132HL0, 43133HL8, 43134HL6, 43135HL3, 43201HL3, 43202HL1, 43203HL9, 44605HL4, 44606HL2, 44607HL0, 44608HL8, 44609HL6, 44610HL4, 44681HL5, 45046HL0, 45047HL8, 45048HL6, 45128HL6, 45129HL4, 45130HL2, 45131HL0, 48119HL2, 48120HL0, 48121HL8, 48122HL6, 48123HL4, 48124HL2, 48125HL9, 48126HL7, 48127HL5, 48128HL3, 48164HL8, 48181HL2, 48274HL5, 48276HL0, 48277HL8, 48278HL6, 48680HL3, 48682HL9, 48683HL7, 48684HL5, 48685HL2, 48686HL0, 48687HL8, 48688HL6, 48689HL4, 48876HL7, 48877HL5, 48878HL3, 48995HL3, 48997HL1, 48998HL9, 49000HL3, 49208HL2, 49209HL0, 49210HL8, 49211HL6, 49391HL6, 49392HL4, 49393HL2, 49395HL7, 49494HL8, 49496HL3, 49497HL1, 49498HL9, 49499HL7, 49500HL2, 49501HL0, 49502HL8, 49503HL6, 49681HL0, 49682HL8, 49683HL6, 49684HL4, 49685HL1, 49686HL9, 49687HL7, 49688HL5, 49850HL1, 49851HL9, 49852HL7, 49853HL5, 49855HL0, 49856HL8, 49857HL6, 50027HL2, 50028HL0, 50029HL8, 50030HL6, 50031HL4, 50177HL5, 50178HL3, 50179HL1, 50180HL9, 50321HL9, 50322HL7, 50323HL5, 50324HL3, 50444HL9, 50445HL6, 50446HL4, 50447HL2, 50448HL0, 50449HL8, 50450HL6, 50451HL4, 50452HL2, 50453HL0, 50790HL5, 50791HL3, 50792HL1, 50793HL9, 50794HL9, 50795HL4, 50797HL0, 50798HL8, 50801HL0, 51143HL6, 51144HL4, 51145HL1, 51147HL7, 51486HL9, 51487HL7, 51488HL5, 52014HL8, and 70RD05001

Root Cause

Radiation Control for Health and Safety Act

Action Taken

GE will be correcting the units by replacing all collimators that don't meet the 160-lux requirement. The field correction will be initiated and is scheduled to be completed by October 2008.