RecallDepth

AST R1 Reagent, 120 ml., Pointe Scientific, Inc., Canton, MI; Catalog no. 8-A7561-R1-120. For the quantitative determination of aspartate amoinotransferase in human serum

This recall has been terminated (originally issued October 22, 2008).

Company
Pointe Scientific, Inc.
Recall Initiated
August 25, 2008
Posted
October 22, 2008
Terminated
November 27, 2009
Recall Number
Z-0172-2009
Quantity
25 vials
Firm Location
Canton, MI
Official Source
View on FDA website ↗

Reason for Recall

The R1 component may be contaminated with Serratia liquefaciens, resulting decreased absorbances and failure of the reagent to produce test results.

Distribution

Nationwide, Bolivia, Ecuador, Greece, Indonesia, Mexico, Nepal, Philippines, Romania and Serbia.

Lot / Code Info

Lot 728504-079; Exp. 4/09.

Root Cause

Material/Component Contamination

Action Taken

Consignees were notified via recall letter dated 8/22/08 and mailed 8/25/08 to examine their inventory, dispose of any recalled product found, and to notify the firm for replacements. Contact Pointe Scientific Inc., at 1-800-757-5313 for assistance.

More recalls by Pointe Scientific, Inc.

View all recalls by this company →