EvoTech Endoscope Cleaner & Reprocessor, Product Code 50004 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic detergent, CIDEZYME GI, to achieve cleaning and high level disinfection of heat sensitive semi-critical endoscopes.

This recall has been terminated (originally issued August 8, 2008).

Company
Advanced Sterilization Products
Recall Initiated
February 13, 2008
Posted
August 8, 2008
Terminated
November 29, 2011
Recall Number
Z-1368-2008
Quantity
64 units
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Cleaning cycles are being cancelled. Specific issues are: a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging, b) Interface board malfunction resulting in a premature lid opening, and c) Minor fluid leaks within the system (commonly at the 0.2 micron filter attachment points).

Distribution

Worldwide Distribution --- including USA and countries of Canada and Japan.

Lot / Code Info

Product Code 50004; Serial Numbers: 50004001070080, 50004001070082, 50004001070083, 50004001070043, 50004001070044, 50004001070046, 50004001070048, 50004001070023, 50004001070024, 50004001070021, 50004001070075, 50004001070045, 50004001070078, 50004001070079, 50004001070107, 50004001070110, 50004001070081, 50004001070062, 50004001070064, 50004001070063, 50004001070034, 50004001070025, 50004001070029, 50004001070047, 50004001070049, 50004001070061, 50004001070087, 50004001070088, 50004001070084, 50004001070077, 50004001070076, 50004001070085, 50004001070086, 50004001070090, 50004001070091, 50004001070094, 50004001070089, 50004001070092, 50004001070098, 50004001070097, 50004001070099, 50004001070100, 50004001070101, 50004001070105, 50004001070106, 50004001070108, 50004001070109, 50004001070102, 50004001070103, 50004001070104, 50004001070020, 50004001070031, 50004001070035, 50004001070036, 50004001070039, 50004001070042, 50004001070059, 50004001070065, 50004001070071, 50004001070072, 50004001070073, 50004001070074, 50004001070114

Root Cause

Other

Action Taken

Feburary 13, 2008 ASP issued an Urgent Product Recall Letter. ASP recommends that customers use an alternative reprocessing method until they return the system. However, if they do not have an alternate method for processing endoscopes and choose to continue using the EvoTech ECR System during this time, endoscopes processed through completed cycles in accordance with the User's Guide will be appropriately cleaned and highlevel disinfected. ASP is prepared to tailor an alternative solution to minimize the interruption to the customer's processing needs.