AxSYM Rubella IgG Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; list 3B23-20 The AxSYM Rubella IgG assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative and qualitative measurement of IgG antibodies to rubella virus in human serum or plasma (EDTA, heparin or sodium citrate) to aid in the determination of immune status to rubella.

This recall has been terminated (originally issued September 16, 2008).

Company
Abbott Laboratories
Recall Initiated
June 2, 2008
Posted
September 16, 2008
Terminated
November 5, 2008
Recall Number
Z-2175-2008
Quantity
33,485 reagent kits
Firm Location
Abbott Park, IL
Official Source
View on FDA website ↗

Reason for Recall

An increase in complaints was noted for AxSYM Rubella IgG assay calibration failures due to error codes related to elevated Calibrator A (or Master Calibrator 1) rates too high.

Distribution

Worldwide Distribution including USA, territory of Puerto Rico, and countries of Australia, Bahamas, Canada, Chile, Colombia, Costa Rica, Curacao, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Jamaica, Mexico, Peru, New Zealand, Panama, Singapore, South Korea, Thailand, Turks & Caicos and Uruguay.

Lot / Code Info

list 3B23-20: lots 61151M101 (exp. 12/28/08), 61151M100 (exp. 12/28/08), 60943M101 (exp. 09/22/08), 60943M100 (exp. 09/22/08), 58296M101 (exp. 10/05/08), 58296M100 (exp. 10/05/08), 58295M101 (exp. 09/22/08), 58295M100 (exp. 09/22/08), 58294M101 (exp. 09/01/08), 58294M100 (exp. 09/01/08), 57577M201 (exp. 08/10/08), 57577M200 (exp. 08/10/08), 56659M201 (exp. 07/21/08), 56659M200 (exp. 07/21/08), 56323M101 (exp. 06/30/08), 56323M100 (exp. 06/30/08), 55922M100 (exp. 06/20/08), 55922M101 (exp. 06/20/08), 55921M101 (exp. 06/09/08), and 55921M100 (exp. 06/09/08)

Root Cause

Device Design

Action Taken

Consignees were sent a Product Correction-Immediate Action Required letter on 6/2/08. The letter informed users of calibration failures due to elevates Calibrator A or Master Calibrator 1 rates when attempting to calibrate the AxSYM Rubella IgG assay and that they must clarify the Rubella IgG Calibrator A or the Rubella IgG Master Calibrator 1 and the Rubella IgG Negative Control by centrifugation at >/= 10,000 x g for 10 minutes prior to testing. Users were requested to return a customer reply letter acknowledging receipt of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.