RecallDepth

Zimmer Patient Helper Reinforced Overhead Bar for Hill-Rom beds with removable headboards; Zimmer Inc., Warsaw, IN; Prod. Nos. 00-0965-000-00 and 00-0965-001-00. The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed.

This recall has been terminated (originally issued September 16, 2008).

Company
Zimmer Inc.
Recall Initiated
July 7, 2008
Posted
September 16, 2008
Terminated
November 12, 2009
Recall Number
Z-2296-2008
Quantity
645 of all systems
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.

Distribution

Nationwide and Canada.

Lot / Code Info

All units shipped between 7/11/07 and 5/19/08.

Root Cause

Device Design

Action Taken

Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.