Simplastin HTF Reagents, 20ml HO, ( and 6 ml HO) bioMerieux, Inc., Durham, NC 27704

This recall has been terminated (originally issued May 18, 2005).

Company
Biomerieux, Inc.
Recall Initiated
March 7, 2005
Posted
May 18, 2005
Terminated
October 16, 2006
Recall Number
Z-0843-05
Quantity
19748 kits
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

The ISI labeling for Simplastin HTF on the Coag-a-Mate MAX and Thrombolyzer instrument platforms is not correct. The Simplastin HTF ISI assignment was mis-assigned and the published values are higher than expected for the listed instrument platforms.

Distribution

Nationwide, Canada, Colombia, India, Philippines, Spain, Thailand

Lot / Code Info

Product numbers 259846, 259847, Lot numbers: 161849, 161800 and 161798. Recall Expanded (03/23/2005) to include these additional lot numbers: 161655, 161710, 161741, 161738, 161742, 1621739, 161743, 161740, 161764, 161763, 161797 and 161799.

Root Cause

Other

Action Taken

Consignees were notified by letter on/about March 7, 2005.