RecallDepth

Synthes Matrix 5.5.mm Left and Right Coronal Bender Product Usage: The Matrix 5.5.mm Left and Right Coronal Bender are used together to bend the 5.5 Matrix Rods in the coronal plane, and is intended for stabilization of the thoracic, lumbar, sacral, and/or iliac spine through screw and/or hook and rod fixation.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
June 22, 2013
Posted
November 26, 2013
Terminated
September 10, 2015
Recall Number
Z-0391-2014
Quantity
209
Firm Location
West Chester, PA

Reason for Recall

A complaint was received of the tips of the Matrix 5.5.mm Left and Right Coronal Bender breaking while bending a spinal rod component of the Matrix System.

Distribution

Worldwide Distribution - USA (nationwide) including GA, FL, CA, UT, IA, AZ, OH and Internationally to Canada.

Lot / Code Info

Parts 03.632.040 and 03.632.041 with various lot nos.  Item Lot Number Number   03.632.040 T935338 03.632.040 T939100 03.632.040 T945062 03.632.040 T953982 03.632.040 T956310 03.632.040 T959990 03.632.040 T959995     Item Lot Number Number   03.632.041 T935343 03.632.041 T939105 03.632.041 T939755 03.632.041 T945067 03.632.041 T952750 03.632.041 T956315 03.632.041 T960000 03.632.041 T960005

Root Cause

Component design/selection

Action Taken

Synthes sent an Urgent Notice: Medical Device Recall letter dated June 22, 2013 to users and sales reps. The letter identified the affected product, problem and actions to be taken. For questions call 610-719-5450.

More recalls by Synthes Usa (Hq), Inc.