Synthes Spiked Washers 13.5/6.5 Spiked Washers 13.5/6.5 are intended to be used in ligament reattachment or fixation, specifically readaptation of torn or avulsed ligaments.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
October 3, 2013
Posted
November 26, 2013
Terminated
December 18, 2015
Recall Number
Z-0394-2014
Quantity
3087
Firm Location
West Chester, PA

Reason for Recall

During a retrospective assessment, Spiked Washers 13.5/6.5 were discovered to be mislabeled as Spiked Washers 13.5/5.5.

Distribution

Nationwide Distribution

Lot / Code Info

part 219.951 with multiple lot#:   2335941 2339634 2397429 2405812 2440475 2442736 2496414 5661855 5663841

Root Cause

Finished device change control

Action Taken

SYNTHES sent an Urgent Notice: Medical Device Recall letter dated October 3, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the Verification Section at the end of the letter by checking the appropriate box indicating affected product located. Return the Verification Form to: Credit/Returns Synthes 1101 Synthes Avenue, Monument, CO 80132 Customers were also asked to return the Verification Form indicating they did not have the affected product also. All documents were to be returned to Synthes by: Fax 1-877-270-5855 Scan/email: FieldAction@synthes.com Customers with questions were instructed to call 610-719-5450 or contact their Synthes Sales Consultant For questions regarding this recall call 610-719-5000..

More recalls by Synthes Usa (Hq), Inc.

View all recalls by this company →