Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only Product Usage: The Integra kits which contain a BD LOR syringe are intended to provide licensed medical professionals with a kit containing drugs and other medical devices that are used to support the administration of general, local or spinal anesthetics. The BD LOR syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.

This recall has been terminated (originally issued December 17, 2013).

Company
Integra Lifesciences Corp. D.B.A. Integra Pain Management
Recall Initiated
October 28, 2013
Posted
December 17, 2013
Terminated
January 29, 2014
Recall Number
Z-0533-2014
Quantity
359,250
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Integra LifeSciences Corporation is recalling certain lots of Integra Kits for Epidural Anesthesia Administration as a result of a Product Advisory Notice from BD Medical for the LOR syringes indicating the plunger of the syringe may stall or stick when travelling within the syringe.

Distribution

USA Nationwide Distribution and one foreign consignee in Mursalat.

Lot / Code Info

Kit Cat No: 10-2433,13-2595,13-2616,13-2628,3104195,3403061,3403150,3403316,3403579,3403727,3403776,3403813,3403989,3404131,3404269,3404293,3404417,3404468,3404578,3503559,CUS025,CUS1121-01,CUS1164-02,CUS1202-04,CUS1298-03,CUS1376,CUS1376,CUS1400-01,CUS1406-03,CUS1415-04,CUS1568-01,CUS1664-01,CUS1696,CUS1723,CUS1760,CUS1773-01,CUS1797,CUS279-02,CUS416,CUS597,CUS628-02,CUS755,PISLL7,RSM394,RSM395,SS1246,SS2036-01,

Root Cause

Component design/selection

Action Taken

BD Medical sent a Product Advisory Notice letter dated August 23, 2013 to Iaffected consignees by via traceable courier. The letter identified the affected product, problem and actions to be taken. Distributor have been asked to forward copies of the notification to their customers. For questions call BD Customer Service at 1-888-237-2762.