Synthes Application Instrument Sternal Zip Flex System intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion

Company
Synthes Usa (Hq), Inc.
Recall Initiated
August 13, 2013
Posted
October 30, 2013
Terminated
August 18, 2016
Recall Number
Z-0114-2014
Quantity
319
Firm Location
West Chester, PA

Reason for Recall

When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.

Distribution

Worldwide Distribution - US (nationwide) and country of Mexico.

Lot / Code Info

Part No.: 03.501.080 with multiple lot numbers: 3783492 3783913 3788496 3822332 7516728 7606881 7635218 7653178 7659168 7666085 7671934 7679825 7689244 7694377 7700691 7705157 7720599 7738572 7738573 7740498 7742713 7767497 7797648 7803768 7806881 7818677 7818682 7821672 7827088 7831855 7833606 7858407 8068078

Root Cause

Labeling False and Misleading

Action Taken

Synthes sent a Urgent Medical Device Notification dated August 13, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken: This notification is to make you aware of the issue and the importance of completely releasing the tensioning trigger before cutting the implant. Please take the following action: "To ensure your receipt and understanding of this information, Please complete the verification section of this letter and return to Synthes by either of the following methods: " Fax: 1-888-352-9616 or " Scan/email: FieldAction@synthes.com If you have any questions, please call 610-719-5450 or contact your Synthes Sales Consultant.