RecallDepth

ACCU-CHECK FastClix Lancing Device. It is intended for use in conjunction with a blood glucose monitoring system.

This recall has been terminated (originally issued August 28, 2013).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
August 19, 2013
Posted
August 28, 2013
Terminated
May 12, 2014
Recall Number
Z-2109-2013
Quantity
36,332
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Internal Roche inspections have revealed that after the lancet is used for a finger stick, the needle may not fully retract back within the lancet drum or back into the cap. This allows the needle to protrude inside the cap or outside of the cap of the ACCU-CHEK FastClix lancing device.

Distribution

Nationwide Distribution

Lot / Code Info

Lot Number(s): GDA 048, GDA050, GDA 051, GDA 063

Root Cause

Equipment maintenance

Action Taken

On 8/19/ 2013, Roche sent Distributor/Retailer URGENT MEDICAL DEVICE CORRECTION notifications which described the recall issue and health consequences, provided product description, and provided actions required of consignees. Roche sent a CUSTOMER BULLETIN to their direct customers. URGENT MEDICAL DEVICE CORRECTION faxback form for consignee acknowledgement of recall is to be faxed to 1-888-240-2804. Consignees may contact the ACCU-CHEK Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 if they have questions about the information contained in the Distributor/Retailer Letter.

More recalls by Roche Diagnostics Operations, Inc.

View all recalls by this company →