RecallDepth

TWINFIX" Ultra 4.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID") with Needles; Part Number: 72202612 Biodegradable suture anchor

This recall has been terminated (originally issued September 17, 2013).

Company
Smith & Nephew, Inc. Endoscopy Division
Recall Initiated
August 6, 2013
Posted
September 17, 2013
Terminated
April 20, 2016
Recall Number
Z-2238-2013
Quantity
1571 US
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.

Distribution

Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.

Lot / Code Info

50373723 50389654 50389655 50397577 50397596 50397599 50397605 50417226 50431744 50431853 50433695 50437528 50440608 50441690 50447116 50447118 50447119 50448227 50454009 50454867 50461412 50467503 50469159 50470653

Root Cause

Packaging

Action Taken

Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.

More recalls by Smith & Nephew, Inc. Endoscopy Division

View all recalls by this company →