Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implantable components of the SynchroMed II infusion system include the pump, catheter,r and catheter accessories The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of drugs or fluids:

This recall has been terminated (originally issued June 25, 2013).

Company
Medtronic Neuromodulation
Recall Initiated
June 3, 2013
Posted
June 25, 2013
Terminated
November 16, 2015
Recall Number
Z-1579-2013
Quantity
261,109 total devices estimated implanted Worldwide
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und

Distribution

worldwide

Lot / Code Info

all SynchroMed II pumps

Root Cause

Device Design

Action Taken

Medtronic sent an Urgent Medical Device Correction letter dated May 2013, to all affected customers. The letter described the issue of the pump's potentially shorting, how to identify if a pump is affected, and provided ongoing patient management recommendations. The Medtronic Rep would complete a Physician Notification Confirmation Form. A copy of the recall notification letter and the completed Confirmation Form is to be left with the physician. Customers with questions were instructed to call 1-800-707-0933. For questions regarding this recall call 763-526-9738.