Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

This recall has been terminated (originally issued March 11, 2013).

Company
Draeger Medical, Inc.
Recall Initiated
December 13, 2012
Posted
March 11, 2013
Terminated
May 20, 2013
Recall Number
Z-0939-2013
Quantity
260
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Distribution

Distributed in the states of CA, IL, and MA.

Lot / Code Info

Catalog Number MS20783 (Infinity MCable-Dual Hemo)

Root Cause

Labeling design

Action Taken

Drager sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-543-5047 for questions relating to this notice.