Vial Label: 1) LIATEST VWF:Ag Buffer - 5 ml Diagnostica Stago 2) LIATEST VWF:Ag Latex + 4 ml R3 Diagnostica Stago 3) LIATEST VWF:Ag Latex Diluent 4 mL Diagnostica Stago Model, catalogue, number 00518 - 58978 - 58979 Kit Label: LIATEST VWF:Ag DIAGNOSTICA STAGO S.A.S. 92600 Asnires sur Seine (France) Coagulation Assay

Company
Diagnostica Stago, Inc.
Recall Initiated
January 28, 2013
Posted
March 14, 2013
Terminated
July 19, 2013
Recall Number
Z-0951-2013
Quantity
2901 kits
Firm Location
Parsippany, NJ

Reason for Recall

Diagnostica Stago has confirmed that sometimes the delta OD measured can be changed leading to a lowered percentage result.

Distribution

Worldwide Distribution - USA and Canada.

Lot / Code Info

Lot/Unit Numbers : Combination of All STA-R¿/ STA-R EVOLUTION¿ and lots of STA¿ - LIATEST¿ VWF: AG are involved.    Lot Number Expiry date 108128 2013-03 108512 2013-04 108778 2013-07 109415 2013-10 109591 2014-01

Root Cause

Software design (manufacturing process)

Action Taken

Diagnostica Stago Inc. sent e-mail notifications on 1/28/2013 to all consignees (laboratory supervisor). The written field safety notice, entitled Important Field Safety Notice, with its response form was attached. The letter identified the affected product and gave a description of the defect. Customers were asked to confirm all the results obtained in the abnormal range by either of the two methods provided. Users are to complete and return the enclosed response form as soon as possible to the Stago Hotline via fax, 973-644-9348. Questions should be directed to the Stago Hotline at 800-725-0607.