ACCU-CHEK Aviva Nano Blood Glucose Meter No product labeling, beyond the labels applied to the meter, was provided with the meter as these meters were never intended to be distributed. A representative copy of product labeling that would have been included in a finished device kit is attached. The Accu-Chek Aviva Nano Blood Glucose meter is intended to be used with Accu-Chek Aviva test strips for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm. The system is intended to be used by a single person and should not be shared. The system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The system should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

This recall has been terminated (originally issued April 9, 2012).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
March 12, 2012
Posted
April 9, 2012
Terminated
November 6, 2012
Recall Number
Z-1386-2012
Quantity
82 meter distributed on Ebay
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Accu-Chek Aviva Nano Blood Glucose meters that were manufactured for the purpose of validating a manufacturing process for product to be distributed outside the US were distributed in the US through eBay by an unauthorized third party. These meters were not intended for distribution, were not verified to meet final product specifications and were not distributed to customers with the necessary pr

Distribution

Nationwide Distribution-USA (nationwide) including the states of AZ, NY, RI, TN, NJ, IN, TX, AL, NV, CA, OR, KY, IL, AR, NM, NE, MD, MO, GA, and SD.

Lot / Code Info

05075572002, 05075599002, 05075602002, 05075629002, Part numbers: 05180848002, 05075246002, 05075530002, 05075548002, 05075556002, 05075564002 these are not finished device part numbers and were not intended for use in the US.

Root Cause

Employee error

Action Taken

Roche sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 14, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to immediately return their ACCU-CHEK Aviva Nano Meter to the firm in an enclosed pre-paid envelope via USPS. Contact Roche Product Removal Services at 1-888-345-5352 between 8:00 am and 5:00 pm for questions regarding this notice.