Disposable Surgical Cable, Product Code: FL-601-97 (100), Single use, Rx Only, Remington Medical, Inc., 6830 Meadowridge Court, Alpharetta, GA 30005 Usage: Temporary pacing of open hearts: used with External Pacing Generators (EPGs).

This recall has been terminated (originally issued August 13, 2010).

Company
Remington Medical Inc.
Recall Initiated
May 19, 2010
Posted
August 13, 2010
Terminated
November 29, 2010
Recall Number
Z-2223-2010
Quantity
1488 units
Firm Location
Alpharetta, GA
Official Source
View on FDA website ↗

Reason for Recall

The Safe Connect component may separate from the cable, causing the cable to become unplugged from the EPG.

Distribution

AZ, CA, CO, CT, DE, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NM, NY, OH, OK, OR, PA, SC, TX, TN, TX, VA, WA, WI and Canada.

Lot / Code Info

Lot number: 100701

Root Cause

Process design

Action Taken

Consignees were notified on/about 05/19/2010. They were instructed to examine their inventory and quarantine product subject to the recall. If they distributed the product, they were advised to identify their customers and notify them of the recall (via the recall letter). A Return Authorization was provided. More information can be obtained by calling 770.888.8520.