Medtronic HMS PLUS, Hemostasis Management System, Model 30514. It is a microprocessor based, multichannel clot timing instrument with automated syringe handling for pipetting blood into single use cartridges. It performs in vitro heparin sensitivity evaluations, heparin assays, activated clotting times and platelet function evaluations.

Company
Medtronic Inc
Recall Initiated
December 14, 2009
Posted
March 11, 2010
Terminated
December 24, 2011
Recall Number
Z-1070-2010
Quantity
580
Firm Location
Minneapolis, MN

Reason for Recall

Recently the US Food and Drug Administration (FDA) issued an alert, communicating a change in the United States Pharmacopeia (USP) monograph for heparin, introducing a new USP reference standard that will in effect, change potency, and harmonize the USP unit dose with the WHO International Standard (IU) unit dose. Users of the Medtronic Hemostasis Management System (HMS Plus) must be aware o

Distribution

All states in USA (Including DC) except RI and VT.

Lot / Code Info

Catalogue Number 30514

Root Cause

Other

Action Taken

Consignees were sent a Medtronic "Urgent Medical Device Notice' dated December 14, 2009. The letter was addressed to "Valued Customer". The letter described the product and the problem. Advised customers to configure the HMS Plus appropriately. The Field Notification was sent to all current HMS Plus users. The current HMS Plus users were identified by the sales of HMS Plus disposable product -Heparin Assay Cartridges from the period of August 1, 2008 to date. This time was considered appropriate, being more than twice the longest expiration date of any of the HMS disposable.

More recalls by Medtronic Inc

View all recalls by this company →