Steris QFC1728 and QFC1729 Quick Connects used in connections and processing endoscopes. The QFC1728 is used to process the Fujinon 250/270/450/470/490 Group 5 and 530/590 series GI Endoscopes w/o Water Jet. The QFC1729 is used to process 250/450 Group 5 and 530/590 series Endoscopes w/ Water Jet.

Company
Steris Corp
Recall Initiated
February 23, 2007
Posted
June 21, 2007
Terminated
March 1, 2012
Recall Number
Z-0967-2007
Quantity
146
Firm Location
Mentor, OH

Reason for Recall

The center insert of the adaptor #6 restrictor could become loose and/or come out of the housing and may affect the sterilization process of the attached endoscope

Distribution

Worldwide-The product was distributed to USA domestic consignees located in MT, CA, NJ, IA, MA, WY, SC, MO, SC, TN, ID, MT, CO, PA, NV, and TX, and to foreign consignees located in Great Britian and Qatar.

Lot / Code Info

Lot numbers: 2914737, 2876316, 2871200, 2876274, 2914752, 2876704, 2876720, 2876746, 2876761, 3072030, and 3072014.

Root Cause

Process control

Action Taken

On February 22, 2007, customers were contacted by phone and were notified by recall letter dated February 23, 2007. The recall notifications indicate that Steris Account managers will visit each consignee and replace affect product.