RecallDepth

ConMed Eliminator Pyloric/Colonic PET Balloon Dilator; Outer Diameter (OD) size 20mm (60 Fr), length 4cm; product code/end item #000861 and #000867 (5 pack); Single Use, Non-Sterile; distributed by ConMed Endoscopic Technologies, Billerica, MA 01821.

This recall has been terminated (originally issued June 19, 2007).

Company
Conmed Endoscopic Technologies, Inc.
Recall Initiated
April 16, 2007
Posted
June 19, 2007
Terminated
July 30, 2010
Recall Number
Z-0949-2007
Quantity
239 units (#000861) and 12 units (#000867).
Firm Location
Billerica, MA
Official Source
View on FDA website ↗

Reason for Recall

Mislabeling issues including; incorrect balloon diameters, lot number discrepancies, and incorrect type of dilator.

Distribution

Worldwide-Devices distributed to medical facilities and distributors Nationwide (USA) and to distributors in the following countries: Australia, Belgium, Canada, Croatia, Denmark, Germany, India, Italy, Japan, Mexico, Poland, Spain and Sweden.

Lot / Code Info

All lots manufactured from May 03, 2006 to December 14, 2006 which includes lot numbers/codes from 0605031 to 0612141. Lot code identification is as follows: For lot #0605031: 06 = year (2006); 05 = month (May); 03 = Day of Month (3rd); 1 = Manufacturing Shift Code (1st shift).

Root Cause

Other

Action Taken

Notification Letters were sent to domestic consignees on 04/16/07 and to international consignees on 04/17/07. Consignees instructed to discontinue product use, sub recall to accounts sold product, and to contact Stericycle, Inc. at 1-800-668-4391 for instructions on returning product.