Stat Profile Critical Care (CCX) Analyzer Catalog Number: 35942

Company
Nova Biomedical Corporation
Recall Initiated
January 27, 2006
Posted
February 28, 2006
Terminated
October 19, 2006
Recall Number
Z-0576-06
Quantity
498 units
Firm Location
Waltham, MA

Reason for Recall

Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs

Distribution

Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan, Greece, Thailand, Israel, Spain, China, Hong Kong, Argentina, Czech Republic, Croatia,Trinidad WI

Lot / Code Info

All S/N''s

Root Cause

Other

Action Taken

Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.