Cypher MIS Screw System-The system includes screws, various types and sizes of rods, and set screws. Various instruments are also available for use by the surgeon to facilitate implantation of the device. Non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion.

This recall has been terminated (originally issued January 5, 2015).

Company
Ebi, Llc
Recall Initiated
October 8, 2014
Posted
January 5, 2015
Terminated
December 21, 2015
Recall Number
Z-0909-2015
Quantity
2816 approx.units ( 2688 units US and 128 OUS)
Firm Location
Parsippany, NJ
Official Source
View on FDA website ↗

Reason for Recall

Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.

Distribution

Worldwide Distribution: US Distribution to states of: AL, CA, CT, IN, LA, NC, NJ TX & WI; and the country of: The Netherlands.

Lot / Code Info

Item Number: 14-571845 5.5 TI 10.5X45MM CANN TRANS 14-571850 5.5 TI 10.5X50MM CANN TRANS 14-571855 5.5 TI 10.5X55MM CANN TRANS 14-571860 5.5 TI 10.5X60MM CANN TRANS 14-571865 5.5 TI 10.5X65MM CANN TRANS 14-571870 5.5 TI 10.5X70MM CANN TRANS 14-571875 5.5 TI 10.5X75MM CANN TRANS 14-571880 5.5 TI 10.5X80MM CANN TRANS 14-571885 5.5 TI 10.5X85MM CANN TRANS 14-571890 5.5 TI 10.5X90MM CANN TRANS 14-571895 5.5 TI 10.5X95MM CANN TRANS 14-571899 5.5 TI 10.5X100MM CANN TRANS

Root Cause

Device Design

Action Taken

The firm, EBI LLC/Biomet, sent an "URGENT PRODUCT INFORMATION" letter dated 10/6/2014 via Fed Ex to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to follow the steps outlined in the letter. If you have any questions, please feel free to contact Director of Regulatory Affairs at 303-501-8548.