Ultradent Products, Chromaclone PVS, Fast Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material, Part No. 5013, Lot No. B7DS3, Used in prosthodontics.

Company
Ultradent Products, Inc.
Recall Initiated
February 4, 2013
Posted
February 13, 2013
Terminated
May 30, 2013
Recall Number
Z-0812-2013
Quantity
100 cartridges
Firm Location
South Jordan, UT

Reason for Recall

Incorrect packaging of regular set and fast set cartridges.

Distribution

Nationwide Distribution

Lot / Code Info

Lot No. B7DS3

Root Cause

Packaging process control

Action Taken

Ultradent Products, Inc. sent an Urgent Medical Device Correction letter dated February 4, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected products. They may continue to use product or dispose unwanted cartridges and reply back to firm indicating acknowledgement of letter and if replacement is needed. Consumers with questions can call the firm at 801-553-4420. For questions regarding this recall call 801-553-4420.