Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

This recall has been terminated (originally issued December 21, 2012).

Company
Radiometer America Inc
Recall Initiated
June 6, 2006
Posted
December 21, 2012
Terminated
February 5, 2013
Recall Number
Z-0608-2013
Quantity
840 Ampules
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Distribution

Distributed to KY, NC, and NY.

Lot / Code Info

Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.

Root Cause

Process control

Action Taken

Radiometer sent Field Action Notes regarding the affected product. The note identified the product, the problem and the action to be taken by the customer. Instead of scanning the barcorde on the insert to enter the QC data into the anyalyzer, customers were instructed to manually key in the QUALICHECK type, lot number, and correct control range. For questions regarding this recall call 440-871-8900.