OEC InstaTrak 3500, picture archiving and communications system intended as an aid to locate anatomical structures during open percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgical technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.
- Company
- Ge Oec Medical Systems, Inc
- Recall Initiated
- October 11, 2006
- Posted
- August 22, 2012
- Terminated
- August 22, 2012
- Recall Number
- Z-2236-2012
- Quantity
- 99 units
- Firm Location
- Salt Lake City, UT
Reason for Recall
GE Healthcare had recalled certain OEC InstaTrak 3500 Carts due to the potential for the cart to tip over when the arm of the imaging device is extended during use.
Distribution
Nationwide Distribution.
Lot / Code Info
Model number A857680 with various serial numbers (refer to consignee list).
Root Cause
Component design/selection
Action Taken
GE Healthcare sent an "URGENT SAFETY ADVISORY NOTICE" dated October 11, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter stated that GEHC OEC Field Service Engineers would be contacting their facility to place a caution label on the articulating monitor arm of the cart and to provide a revised operators manual that will contain additional information on the safe usage of the cart. Until a GEHC Engineer had taken the actions described, they advised their consignees to exercise caution when fully extending the articulating monitor arm on the device to avoid potentially tipping over the cart. Contact the firm at 800-874-7378 for questions regarding this notice.