OEC InstaTrak 3500, picture archiving and communications system intended as an aid to locate anatomical structures during open percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgical technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.

Company
Ge Oec Medical Systems, Inc
Recall Initiated
October 11, 2006
Posted
August 22, 2012
Terminated
August 22, 2012
Recall Number
Z-2236-2012
Quantity
99 units
Firm Location
Salt Lake City, UT

Reason for Recall

GE Healthcare had recalled certain OEC InstaTrak 3500 Carts due to the potential for the cart to tip over when the arm of the imaging device is extended during use.

Distribution

Nationwide Distribution.

Lot / Code Info

Model number A857680 with various serial numbers (refer to consignee list).

Root Cause

Component design/selection

Action Taken

GE Healthcare sent an "URGENT SAFETY ADVISORY NOTICE" dated October 11, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter stated that GEHC OEC Field Service Engineers would be contacting their facility to place a caution label on the articulating monitor arm of the cart and to provide a revised operators manual that will contain additional information on the safe usage of the cart. Until a GEHC Engineer had taken the actions described, they advised their consignees to exercise caution when fully extending the articulating monitor arm on the device to avoid potentially tipping over the cart. Contact the firm at 800-874-7378 for questions regarding this notice.

More recalls by Ge Oec Medical Systems, Inc

View all recalls by this company →