GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. In certain fluoroscope X-ray modes, the cable provides a signal to the auto-injector to trigger a contrast injection. Part numbers include 00-902260-01 and 00-883654-01.

Company
Ge Oec Medical Systems, Inc
Recall Initiated
October 10, 2008
Posted
August 23, 2012
Terminated
January 10, 2013
Recall Number
Z-2262-2012
Quantity
149 units
Firm Location
Salt Lake City, UT

Reason for Recall

GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.

Distribution

Nationwide distribution.

Lot / Code Info

Serial Numbers: 82-1138, 82-1586, 82-2089, 82-2117, 82-2182, 82-2214, 82-2226, 82-2262, 82-2264, 82-2308, 82-2312, 82-2314, 82-2321, 82-2328, 82-2380, 82-2391, 82-2438, 82-2483, 82-2512, 82-2516, 82-2535, 82-2722, 82-2796, 82-2821, 82-2846, 82-2848, 82-3052, 82-3055, 82-3129, 82-3144, 82-3170, 82-3171, 82-3201, 82-3256, 82-3267, 82-3314, 82-3350, 82-3361, 82-3377, 82-3467, 82-3468, 82-3659, 82-3709, 82-3722, 82-3859, 82-7016-MH, 82-7022-MH, 82-7037-MH, 82-7055-MH, 82-7062-MH, 82-7080-MH, 82-7086-MH, 82-7105-MH, 82-7114-MH, 82-7124-MH, 82-7131-MH, 82-7133-MH, 82-7137-MH, 82-7141-MH, 82-7154-CMH, 82-7155-MH, 82-7156-MH, 82-7161-MH, 82-7175-MH, 82-7189-MH, 82-7190-MH, 89-1441, 8S-1053, 8S-1436, 8S-2059-C, 8S-7021-CMH, 8S-7023-CMH, 8S-7040-CMH, 8S-7087-CMH, 8S-7164-CMH, DEMO/8S-1355-N, DEMO/E3-7002-MH, E2-0049, E2-0088, E2-0111, E2-0262, E2-7001-MH, E2-7007-MH, E2-7010-MH, E2-7014-MH, E2-7015-MH, E2-7031-MH, E2-7032-MH, E2-7035-MH, E2-7043-MH, E2-7059-MH, E2-7061-MH, E2-7067-MH, E2-7075-MH, E2-7076-MH, E2-7015-MH, & rENTAL/82-7004-MH.

Root Cause

Device Design

Action Taken

An Urgent Safety Advisory Notice was sent to consignees on October 10, 2008. The letter identified the affected product and notified customers of the situation, along with instructing them to discontinue use until proper validation and instruction update for cable use was completed. GEHC OEC was to have contacted customers when the solution became available and resolved it at no cost to the customer.