Celsus prosthetic foot, Product Number 103658-00. The Celsus is intended for low impact level lower limb prosthesis users weighing up to 300 lbs.

This recall has been terminated (originally issued May 23, 2012).

Company
College Park Industries, Inc.
Recall Initiated
April 5, 2012
Posted
May 23, 2012
Terminated
July 31, 2012
Recall Number
Z-1650-2012
Quantity
11
Firm Location
Fraser, MI
Official Source
View on FDA website ↗

Reason for Recall

An incorrect shorter part ( pyramid) was installed in the foot. A foot of this size and configuration requires a longer pyramid. Customers requiring the extra firm configuration may have a minimal risk of component failure, which could be implicated in a potential fall.

Distribution

Worldwide Distribution - USA, including the states of CA, MA, ME, WA, TX, VA, AZ and the country of England.

Lot / Code Info

Product Number103658-00, Lot/ Serial Number (USA)TE012171, TE012151, TE012142, TE012159, TE012158, TE012154, TE012186, TE012185, TE012161, TE012139, (UK)TE012182.

Root Cause

Mixed-up of materials/components

Action Taken

All customers who received affected product were initially contacted via phone on April 5, 2012 and made aware of the recall, including instructions to notify customers immediately and arrange replacements. A "Recall Notice" letter dated April 24, 2012 was sent to customers describing the problem and the recommended actions. Contact reference person is Kimberly Light, 1 800-728-7950, info@college-park.com.