RecallDepth

DPM 6 Monitor in use with a Multi Parameter Module (MPM). Patient monitor used for monitoring, displaying, reviewing, storing, and alarming of multiple physiological parameters including ECG, arrhythmia detection, ST segment analysis, heart rate (HR), respiration rate (RESP), temperature, pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), cardiac output (C.O.) and respiration mechanics (RM)

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
March 5, 2014
Posted
May 13, 2014
Terminated
April 27, 2015
Recall Number
Z-1611-2014
Quantity
4 units
Firm Location
Mahwah, NJ

Reason for Recall

The MPM included a cardiac ST/Arrhythmia analysis feature which is not cleared for marketing.

Distribution

US Nationwide Distribution in the state of OH

Lot / Code Info

monitor p/n 6802F-PA00001  module p/n 115-001704-00/M51A-30-80876

Root Cause

No Marketing Application

Action Taken

Mindray DS Usa Inc. contacted their sole customer via phone on March 5, 2014. The affected products will receive software upgrade.

More recalls by Mindray Ds Usa, Inc. D.B.A. Mindray North America