Guidant LATITUDE Communicator, model 6481, Boston Scientific 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. The LATITUDE Patient Management system enables physicians to periodically monitor both patient and device status remotely. The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Guidant or Boston Scientific pulse generator and transfer data to a central database.

This recall has been terminated (originally issued November 12, 2009).

Company
Boston Scientific Crm Corp
Recall Initiated
August 7, 2009
Posted
November 12, 2009
Terminated
December 4, 2012
Recall Number
Z-0180-2010
Quantity
97
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

The L ATITUDE communicator (Model 6481) for a patient may not have been able to receive measurements from their weight scale or blood pressure monitor afer initial setup. The issue has been resolved. If any patient weight or blood pressure readings are stored in the patient's sensor, they will be uploaded to LATITUDE after the patient's communicator calls in to Boston Scientific's servers. It

Distribution

Nationwide Distribution -- AZ, AR, CA, CT, FL, GA, IL, IN, MA, MO, NV, NJ, NY, OH, OR, PA, SC, TX, UT, VA and WA.

Lot / Code Info

SERIAL NUMBERS:  029251, 064452, 059718, 064001, 056441, 059851, 070188, 072291, 312706, 305339, 312762, 004610, 024762, 049668, 061205, 067781, 051788, 056069, 068447, 069808, 008653, 068085, 057090, 023576, 028974, 017638, 603254, 064457, 069473, 069902, 071065, 072021, 073621, 073837, 312707, 605519,  603234, 603788, 072396, 059662, 056346, 301645, 312781, 307327, 067067,  055853, 060274, 605304, 305344, 302301, 605981, 009271, 056500, 300895,  602642, 030830, 301117, 055001, 049660, 047455, 056923, 309556, 044034,  063667, 005931, 067765, 063012, 050130, 051267, 055558, 060325, 060973,  063034, 306808, 306852, 309355, 041156, 062922, 063064, 061177, 300499,  039398, 039401, 039403, 060619, 065476, 058913, 056802, 073812  024833, 073560, 606104, 606288, 606604, 606826, 607180 608231 and 607180.

Root Cause

Software design

Action Taken

Boston Scientific used phone calls as the delivery method for these communications. The communication began August 6, 2009 for the original 6 patients affected. Notifications via phone calls to additional patients affected began on August 17, 2009. Direct questions about the recall to the LATITUDE Customer Service by calling 1-651-582-4000.