ARCHITECT i2000 System and ARCHITECT i2000 SR wash buffer and wash zone mechanism grounding kit. ARCHITECT System is designed to perform automated chemistry tests utilizing photometry and potentiometric technology; and immunoassay tests utilizing chemiluminescent microparticle assay detection technology.

This recall has been terminated (originally issued November 9, 2009).

Company
Abbott Laboratories, Inc
Recall Initiated
July 17, 2009
Posted
November 9, 2009
Terminated
January 26, 2010
Recall Number
Z-0132-2010
Quantity
4,526
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

Under certain conditions, the product's wash buffer containing sodium azide can come in contact with the wash zone mechanism ground cables on either of the two wash zones. The copper in the ground strap may corrode and form an unstable chemical substance (metal azide) that may be sensitive to direct pressure and impact.

Distribution

Worldwide Distribution

Lot / Code Info

ARCHITECT i2000 System List Number 8C89-01; ARCHITECT i2000SR System System List Number 3M74-01 and 3M74-02; Wash Zone Mechanism Grounding Kit Part Number 7-201369-01; Concentrated Wash Buffer 1 L List Number 6C54-58 and l0 L List number 6C54-88.

Root Cause

Other

Action Taken

A product correction notice was issued to consignees on July 17, 2009. Field representatives were instructed to visit consignees, inspect cables and replace them