Diasorin LIAISON (R) Chemiluminescence Analyzer Model # 15970. Colorimeter, Photometer, Spectrophotometer for clinical use.

Company
Diasorin Inc.
Recall Initiated
December 5, 2008
Posted
March 30, 2009
Terminated
May 22, 2012
Recall Number
Z-1042-2009
Quantity
9
Firm Location
Stillwater, MN

Reason for Recall

DiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer.

Distribution

Worldwide Distribution -- USA including states of NC, NJ, PA, CA, WA, MO, UT, and OR and country of Canada.

Lot / Code Info

Serial numbers: 2229000828, 2229000829, 2229000839, 2229000857, 2229000859, 2229000926, 2229000972, 2229001196, and 2229001230.

Root Cause

Device Design

Action Taken

DiaSorin Inc authorized Field Service Engineer has visited the sites and performed the necessary field correction. Upon completion of their work, the field service engineer generated a Field Service Report and a copy was left with the customer.