Diasorin LIAISON (R) Chemiluminescence Analyzer Model # 15970. Colorimeter, Photometer, Spectrophotometer for clinical use.
- Company
- Diasorin Inc.
- Recall Initiated
- December 5, 2008
- Posted
- March 30, 2009
- Terminated
- May 22, 2012
- Recall Number
- Z-1042-2009
- Quantity
- 9
- Firm Location
- Stillwater, MN
Reason for Recall
DiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer.
Distribution
Worldwide Distribution -- USA including states of NC, NJ, PA, CA, WA, MO, UT, and OR and country of Canada.
Lot / Code Info
Serial numbers: 2229000828, 2229000829, 2229000839, 2229000857, 2229000859, 2229000926, 2229000972, 2229001196, and 2229001230.
Root Cause
Device Design
Action Taken
DiaSorin Inc authorized Field Service Engineer has visited the sites and performed the necessary field correction. Upon completion of their work, the field service engineer generated a Field Service Report and a copy was left with the customer.