Abbott Clinical Chemistry Creatine Kinase, Model LN 7D63-30, Lot Number: 52044HW00; Expires October 19, 2007, RCR#2018433-10/25/07-006R Abbott Laboratories, Inc., South Pasadena CA 91030

This recall has been terminated (originally issued March 19, 2008).

Company
Abbott Laboratories Inc.
Recall Initiated
October 12, 2007
Posted
March 19, 2008
Terminated
March 28, 2012
Recall Number
Z-1109-2008
Quantity
334 Distributed Worldwide; 164 Domestically
Firm Location
South Pasadena, CA
Official Source
View on FDA website ↗

Reason for Recall

Three potential issues with Creatine Kinase Lot Number: 52044HW00: (1) Decrease in quality control and/or patient results (2) An iIncreased imprecision, or (3) Error Code 1054 (Unable to calculate results, Reaction Failure) on Architect cSystems.

Distribution

Worldwide Distribution --- including USA and countries of Canada, Chili, Uruguay, Germany, Hong Kong, and New Zealand.

Lot / Code Info

Lot Number: 52044HW00

Root Cause

Nonconforming Material/Component

Action Taken

On October 12, 2007, A Product Recall letter was provided to all Clinical Chemistry Creatine Kinase Reagent customers that have received lot number 52044HW00 . Customer reply form were included in all US customers letter. The letter instructed customers to discontinue use of the kits and destroy them following their laboratory procedure, and document the number of kits discarded. Abbott will provide reimbursement for destroyed kits.