RecallDepth

1,25-Dihydroxyvitamin D RIA Kit -For the quantitative determination of 1,25 Dihydroxyvitamin D in serum or EDTA plasma-DiaSorin, Stillwater, Minnesota 55082-0285, USA

Company
Diasorin Inc.
Recall Initiated
June 5, 2006
Posted
April 3, 2007
Terminated
June 25, 2007
Recall Number
Z-0707-2007
Quantity
715
Firm Location
Stillwater, MN

Reason for Recall

Potential for the DiaSorin Kit 1,25-Dihydroxyvitamin D RIA Kit Control 1 (lot 548520) & 2 (lot 548521) to recover out of the defined range, LOW. (If one or both of the kit controls recover outside the defined range, the run is considered invalid)

Distribution

Worldwide Distribution-USA including states of UT, CO, IL, NY, NV, MI, NC, MN, VA, GA, CA, KS, VT, CT and IN, and Countries of Canada, Sweden, Germany, Belgium, Italy, France, Norway, Australia, South Korea, Taiwan, Israel, New Foundland, India, China, Brazil, Austria, Saudi Arabia, and Spain

Lot / Code Info

Catalog #65100E. Kit Lot #''s 117139, 117140, 1117141

Root Cause

Other

Action Taken

Product recalled by letter dated 6/01/2006

More recalls by Diasorin Inc.