GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Information Technologies, Inc., 8200 W. Tower Ave. Milwaukee, Wisconsin USA. (System hardware consists of a processing unit and a monitor)

This recall has been terminated (originally issued June 6, 2007).

Company
General Electric Medical Systems Information Technology
Recall Initiated
January 24, 2007
Posted
June 6, 2007
Terminated
December 10, 2011
Recall Number
Z-0891-2007
Quantity
5,190 Units
Firm Location
Milwaukee, WI
Official Source
View on FDA website ↗

Reason for Recall

Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot .

Distribution

Nationwide

Lot / Code Info

All serial numbers of Software Version 4.1

Root Cause

Other

Action Taken

GE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07 that explained the short and long term solutions and that the company would be providing a software update that will rectify the issue specific to CIC Pro v4.1 software.

More recalls by General Electric Medical Systems Information Technology