GE Discovery VCT, Volume PET-CT scanner, Model 5124069-3.

This recall has been terminated (originally issued May 15, 2007).

Company
Ge Healthcare
Recall Initiated
December 21, 2006
Posted
May 15, 2007
Terminated
July 3, 2012
Recall Number
Z-0806-2007
Quantity
4
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.

Distribution

Worldwide Distribution: USA in the following states and locations: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, Guam and Puerto Rico. Foreign countries include ; Albania, Algeria, Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Canada, Chile, China, Columbia, Czechoslovakia, Denmark, Finland, France, Germany, Great Britain, Greece, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, and Vietnam

Lot / Code Info

All serial numbers

Root Cause

Other

Action Taken

A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.

More recalls by Ge Healthcare

View all recalls by this company →