BacT/VIEW Software, all versions. in vitro diagnostic database software.

This recall has been terminated (originally issued May 18, 2007).

Company
Biomerieux, Inc.
Recall Initiated
December 19, 2006
Posted
May 18, 2007
Terminated
December 8, 2009
Recall Number
Z-0814-2007
Quantity
1233 units
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

BacT/VIEW will be unable to run scheduled macro applications unless the user touches the BacT/VIEW touch screen or clicks a function after resynchronizing the clocks.

Distribution

Worldwide, including USA, Canada, UK, Italy, France, Spain, Germany, Poland, Chile, Japan and the Netherlands.

Lot / Code Info

Product Numbers: Software Version C.30a -22105-68, Update Disk; 22105-69 Install Disk

Root Cause

Other

Action Taken

A customer notification was sent on/about December 19, 2006. Customers will be sent a new BacT/VIEW software version C.30b QTR 1 2007.